 Index 2 SpO2 Simulator
Index 2 SpO2 Simulator
	 
	Index 2 Series combines Fingersense our new patented simulated finger with the first comprehensive probe and electronics test to fully analyze pulse oximeters and probes. 
	 
	BIOTEK Index 2 Series SPO2 Simulator,προσομοιωτή οξύμετρο,محاكاة مقياس التأكسج,oximeter simulator,血氧饱和度模拟器
	 
	OVERVIEW
	 
	Index 2’s new "Fingersense" assures compatibility with virtually any oximeter on the market today. No hardware or firmware upgrades are necessary. Fingersense incorporates new electronic physiology that instantly recognizes and adapts to all types of oximeter simulation technologies. Use its new user definable "make" feature and pulse oximeter makes that are not included within Index can be verified with acceptable accuracy. With the rapid increase of SPO2 monitors, the new physiology in the electronics help insure that Index 2 will be compatible with your pulse oximeter. 
	 
	Electrical simulation for testing the pulse oximeters alone and probe testing is now available in the "E" version. This allows you to test the oximeter without the probe in-line and all simulations are available. The probe analysis includes:
	 
	 
	 
	LED test: Voltage drop across both red and infrared diodes
	 
	Photodiode test: Detector response to a 1 mA illuminating current (both red and infrared)
	 
	Resistive wire check: Resistances between all wires
	 
	Index 2 lets you simulate arterial oxygen saturation (SPO2) from 35% to 100% oxygenation. Heart rate can also be varied from 30 to 250 BPM. In addition, Plethysmographic waveforms can be set from 0-100% amplitude, simulating the range from no pulse up to a very strong pulse. These basic features allow Index 2 to simulate several different patient conditions such as: 
	 
	Normal Patient
	 
	Patient with a weak pulse
	 
	Geriatric patient
	 
	 
	 
	Obese patient
	 
	Bradycardia
	 
	Tachycardia
	 
	Bio-Tek's test simulations verify the pulse oximeters SpO2 and heart rate alarms for response time. Use Index 2 for pre-purchase evaluations to determine how well different pulse oximeters respond to artifacts such as motion, ambient sunlight and 50/60 Hz artificial light. 
	 
	In addition, Index 2 comes with full computer control, a built-in RS232, and a printer port. Test results may be transferred to Bio-Tek's Equipment Management System via OTIS for Windows (OFW) or sent directly to an attached printer for hard-copy results. 
	 
	 
	 
	FEATURES
	 
	Portable
	 
	9 preloaded manufacturers R-Curves
	 
	User definable ‘makes’ for most other manufacturers
	 
	5 downloadable R-curve spaces available >
	 
	Patented physiological finger for a complete SPO2 test
	 
	Electrical probe test and electrical oximeter simulation
	 
	RS232 Port
	 
	 
	 
	APPLICATIONS
	 
	On-going preventative maintenance testing
	 
	Isolate sensor versus electronics problems
	 
	Determine performance variations in less than ideal conditions
	 
	Incoming inspections and repair
	 
	Pre-purchase evaluations
	 
	 
	 
	SPECIFICATIONS
	 
	 
	 
	%O2 
	 
	Range: 35-100%
	 
	Resolution: 1%
	 
	Accuracy: 100-75% ±1%, ±accuracy of the pulse oximeter under test 74-50% ±2% ±accuracy of the pulse oximeter under test Less than 50% unspecified
	 
	Repeatability: ±1 standard deviation
	 
	 
	 
	Rate
	 
	Range: 30-250BPM
	 
	Resolution: 1 BPM
	 
	Accuracy: 1% ±1BPM
	 
	 
	 
	Pulse Amplitude
	 
	Range: 0-100% of nominal pleth amplitude
	 
	Resolution: 1%
	 
	Accuracy: ±1% of setting, nominal pulse
	 
	 
	Display
	 
	2 line x 24 character super twist LCD
	 
	 
	 
	ACCESSORIES
	 
	OTIS for Windows Test Integration Software
	 
	Printer
	 
	Carrying case
	 
	RS232 cable
	 
	Printer cable
	 
	Battery eliminator
	 
	Upgrades
	 
	VERSIONS
	 
	 
	Index 2F
	Optical finger; now with enhanced optics for a more physiological finger
	Index 2E
	Electrical oximeter simulation/testing and electrical probe testing
	Index 2FE
	Both optical finger and electrical simulations/probe testing