Product Library

DefiXpress Defibrillator Technical Specification


Meditech Debrillator Monitor can work both in Automated External Debrillator (AED) mode and Manual mode and record parameters of ECG, HR and SpO2. Features: ·
Works both in automated external debrillator (AED) and Manual modes · Records ECG, HR and SpO2 · It has voice and visual prompts for charge prompt and shockable rhythm · It has a high-resolution display and a thermal printer


Display:
7 screen(17.7cm)diagonally
Sweep speed:
25mm/ sec
Number of leads for ECG:
3-lead ECG cable or 5-lead ECG cable, paddles
Output:
Biphasic
Energy sequence:
Monitor display indicates both selected and delivered energy
Charge time:
<7sec with a new fully charged battery
Charge prompts:
Voice and visual prompts
Electrode impedance measurement range:
0-250Ω
Lead selection:
Displayed on monitor, paddles, I, II, III, AVR, AVL, AVF, V
ECG size:
0.25, 0.5, 1, 1.5, 2, 4 cm/mV display on monitor
Heart rate:
20-300bpm
Heart rate alarm:
On/Off displayed on monitor, user-selectable
Smart alarm:
Beeper/voice prompts indicate shockable rhythm
SpO2 measurement range:
30~100%, ±2% between 80%~90%, others ±5%
Alarm range for SpO2 measurement range:
User set high limit and low limit
Alarm accuracy for SpO2 measurement range:
±10% within setting values
Alarm time accuracy for SpO2 measurement range:
Less than 12secs
Functions of AED mode:
Auto analyse and charge x3 with programmable auto energy level selection, screen prompts, and voice prompts
Shockable rhythms of AED mode:
Ventricular fibrillation with amplitude >=200uV ventricular tachycardia with rates >=140bpm QRS complex wave duration >=140ms
Charge control of AED mode:
Control on device front panel, press key on paddle
Prompts of AED mode:
Voice and visual prompts
Energy selection of Manual mode:
Selectable at 2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, 360 joules
Synchronization of Manual mode:
Synchronizes defibrillator pulse to patient's R-wave. ”SYNC” message displayed on monitor
Printer paper:
50mm thermal paper
Paper speed:
12.5mm/sec, 25mm/sec, 50mm/sec. User-selectable 6-second delay
Printing method:
High-resolution, thermal print head
Print-out modes:
Manual or automatic, user-configurable
On/Off control:
Front panel and paddle
Automatic function:
9-seconds recording initiated by alarm activation or defibrillator charge or defibrillator discharge
Battery:
Rechargeable, Ni-MH battery, 12V
Battery life:
For a new, fully charged battery- 60 defibrillator discharges at maximum energy, or 3 hours minimum of continuous ECG monitoring. Additional parameters will effect operating time with different functions

Promoted Products

Meditech Brands

Meditech Equipment Co.,Ltd is part of Meditech Group. Product(s) described may not be licensed or available for sale in all countries. Sonotech, Sonovet, iSonic, FOs2pro, Dolphi, Defi, HeartRec,miniScan,Cardios,SpirOx,iBreath, Meditech and all corresponding design marks are trademarks of Meditech. The symbol indicates the trademark is registered. Patent and Trademark Office and certain other countries. All other names and marks mentioned are the trade names, trademarks or service marks of their respective owners. Please see the Instructions for Use for a complete listing of the indications, contraindications, warnings and precautions.