ECG
Lead Mode 5 Leads ( R L F N C or RA LA LL RL V)
Lead selection I II III avR avL avF V
Waveform 2 ch
Lead mode 3 Leads ( R L F or RA LA LL)
Lead selection I II III
Waveform 1 ch
Gain ′2.5mm/mV ′5.0mm/mV ′10mm/mV ′20mm/mV
HR and Alarm
Range
Adult 15 ~ 300 bpm
Neo/Ped 15 ~ 350 bpm
Accuracy ±1% or ±1bpm which great
Resolution 1 bpm
Sensitivity > 200 uV P-P
Differential Input Impedance > 5 MΩ
CMRR
Monitor > 105 dB
Operation > 105 dB
Diagnosis > 85 dB
Electrode offset potential ±300mV
Leakage Current < 10 uA
baseline Recovery < 3 s After Defi.
ECG Signal Range ±8 mV (Vp-p)
Bandwidth
Surgery 1 ~ 15 Hz
Monitor 0.5 ~ 35 Hz
Diagnostic 0.05 ~ 100 Hz
Calibration Signal 1 mV (Vp-p) ±5% Accuracy
ST Segment Monitoring Range
Measure and Alarm -2.0 ~ +2.0 mV
ARR Detecting
Type ASYSTOLE VFIB/VTAC COUPLET BIGEMINY TRIGEMINY R ON T VT>2 PVC TACHY BRADY MISSED BEATS PNP PNC
Alarm Available
Review Available
RESPARATION
Method Impedance between R-F(RA-LL)
Differential Input Impedance >2.5 MΩ
Measuring Impedance Range: 0.3~5.0Ω
base line Impedance Range: 0 – 2.5 KΩ
Bandwidth 0.3 ~ 2.5 Hz
Resp. Rate
Measuring and Alarm Range Adult 0 ~ 120 rpm
Neo/Ped 0 ~ 150 rpm
Resolution 1 rpm
Accuracy ±2 rpm
Apean Alarm 10 ~ 40 S
NIBP
Method Oscillometric
Mode Manual Auto STAT
Measuring Interval in AUTO Mode 1 2 3 4 5 10 15 30 60 90 120 180 240 480 Min
Measuring Period in STAT Mode 5 Min
Pulse Rate Range 40 ~ 240 bpm
Alarm Type SYS DIA MEAN
Measuring and alarm Range Adult Mode
SYS 40 ~ 270 mmHg
DIA 10 ~ 215 mmHg
MEAN 20 ~ 235 mmHg
Pediatric Mode
SYS 40 ~ 200 mmHg
DIA 10 ~ 150 mmHg
MEAN 20 ~ 165 mmHg
Neonatal Mode
SYS 40 ~ 135 mmHg
DIA 10 ~ 100 mmHg
MEAN 20 ~ 110 mmHg
Resolution
Pressure 1mmHg
Accuracy Pressure
Maximum Mean error ±5mmHg
Maximum Standard deviation ±8mmHg
Overpressure Protection
Adult Mode 297±3 mmHg
Pediatric Mode 240±3 mmHg
Neonatal Mode 147±3 mmHg
SpO2
Measuring Range 0 ~ 100 %
Alarm Range 0 ~ 100 %
Resolution 1 %
Accuracy 70% ~ 100% ±2 %
0% ~ 69% unspecified
Actualization interval about 1Sec.
Alarm Delay 10 Sec.
Pulse Rate
Measuring and Alarm Range
0~254bpm
Resolution 1bpm
Accuracy ±2bpm
Alarm Delay 10 Sec.
TEMPERATURE
Channel 2
Measuring and Alarm Range 0 ~ 50 °C
Resolution 0.1°C
Accuracy ±0.1°C
Actualization interval about 1 Sec.
Average Time Constant < 10 Sec.
Options: EtCO2 IBP Recorder