Education

BIOTEK Index 2 Series SPO2 Simulator

 

BIOTEK Index 2 Series SPO2 Simulator,προσομοιωτή οξύμετρο,محاكاة مقياس التأكسج,oximeter simulator,血氧饱和度模拟器Index 2 SpO2 Simulator

 
Index 2 Series combines Fingersense our new patented simulated finger with the first comprehensive probe and electronics test to fully analyze pulse oximeters and probes. 
 
BIOTEK Index 2 Series SPO2 Simulator,προσομοιωτή οξύμετρο,محاكاة مقياس التأكسج,oximeter simulator,血氧饱和度模拟器
 
OVERVIEW
 
Index 2’s new "Fingersense" assures compatibility with virtually any oximeter on the market today. No hardware or firmware upgrades are necessary. Fingersense incorporates new electronic physiology that instantly recognizes and adapts to all types of oximeter simulation technologies. Use its new user definable "make" feature and pulse oximeter makes that are not included within Index can be verified with acceptable accuracy. With the rapid increase of SPO2 monitors, the new physiology in the electronics help insure that Index 2 will be compatible with your pulse oximeter. 
 
Electrical simulation for testing the pulse oximeters alone and probe testing is now available in the "E" version. This allows you to test the oximeter without the probe in-line and all simulations are available. The probe analysis includes:
 
 
 
LED test: Voltage drop across both red and infrared diodes
 
Photodiode test: Detector response to a 1 mA illuminating current (both red and infrared)
 
Resistive wire check: Resistances between all wires
 
Index 2 lets you simulate arterial oxygen saturation (SPO2) from 35% to 100% oxygenation. Heart rate can also be varied from 30 to 250 BPM. In addition, Plethysmographic waveforms can be set from 0-100% amplitude, simulating the range from no pulse up to a very strong pulse. These basic features allow Index 2 to simulate several different patient conditions such as: 
 
Normal Patient
 
Patient with a weak pulse
 
Geriatric patient
 
 
 
Obese patient
 
Bradycardia
 
Tachycardia
 
Bio-Tek's test simulations verify the pulse oximeters SpO2 and heart rate alarms for response time. Use Index 2 for pre-purchase evaluations to determine how well different pulse oximeters respond to artifacts such as motion, ambient sunlight and 50/60 Hz artificial light. 
 
In addition, Index 2 comes with full computer control, a built-in RS232, and a printer port. Test results may be transferred to Bio-Tek's Equipment Management System via OTIS for Windows (OFW) or sent directly to an attached printer for hard-copy results. 
 
 
 
FEATURES
 
Portable
 
9 preloaded manufacturers R-Curves
 
User definable ‘makes’ for most other manufacturers
 
5 downloadable R-curve spaces available >
 
Patented physiological finger for a complete SPO2 test
 
Electrical probe test and electrical oximeter simulation
 
RS232 Port
 
 
 
APPLICATIONS
 
On-going preventative maintenance testing
 
Isolate sensor versus electronics problems
 
Determine performance variations in less than ideal conditions
 
Incoming inspections and repair
 
Pre-purchase evaluations
 
 
 
SPECIFICATIONS
 
 
 
%O2 
 
Range: 35-100%
 
Resolution: 1%
 
Accuracy: 100-75% ±1%, ±accuracy of the pulse oximeter under test 74-50% ±2% ±accuracy of the pulse oximeter under test Less than 50% unspecified
 
Repeatability: ±1 standard deviation
 
 
 
Rate
 
Range: 30-250BPM
 
Resolution: 1 BPM
 
Accuracy: 1% ±1BPM
 
 
 
Pulse Amplitude
 
Range: 0-100% of nominal pleth amplitude
 
Resolution: 1%
 
Accuracy: ±1% of setting, nominal pulse
 
 
Display
 
2 line x 24 character super twist LCD
 
 
 
ACCESSORIES
 
OTIS for Windows Test Integration Software
 
Printer
 
Carrying case
 
RS232 cable
 
Printer cable
 
Battery eliminator
 
Upgrades
 
VERSIONS
 
 
Index 2F
Optical finger; now with enhanced optics for a more physiological finger
Index 2E
Electrical oximeter simulation/testing and electrical probe testing
Index 2FE
Both optical finger and electrical simulations/probe testing
 
 
 
 
 
 
 
 
 
 
 
 

Promoted Products

Meditech Brands

Meditech Equipment Co.,Ltd is part of Meditech Group. Product(s) described may not be licensed or available for sale in all countries. Sonotech, Sonovet, iSonic, FOs2pro, Dolphi, Defi, HeartRec,miniScan,Cardios,SpirOx,iBreath, Meditech and all corresponding design marks are trademarks of Meditech. The symbol indicates the trademark is registered. Patent and Trademark Office and certain other countries. All other names and marks mentioned are the trade names, trademarks or service marks of their respective owners. Please see the Instructions for Use for a complete listing of the indications, contraindications, warnings and precautions.

Legal notice Terms and conditions Cookie policy Privacy Policy Professional organisations Careers

2024© Copyright احدى فروع شركة ميديتك جروب و المركز الرئيسي الإقليمي لمنطقتي الشرق الأوسط و شمال إفريقيا | Meditech Group (Middle East branch).

Meditech Group

To Top