• patient monitor 15inch touch screen
  • patient monitor 15inch touch screen
  • patient monitor 15inch touch screen
15 inch patient monitor 192-hour trend graph,1000 group of blood pressure list,ST segment measurements, arrhythmia analysis and more functions,
Support up to 32 TCP/IP network connection,composed of a central information system..
Comes with remote control


Masimo spo2 module (Option)

Product Features

    15” high resolution color TFT display with maximum 13 waveforms
    Full touch-screen enables intuitive operation by clicking on the specific parameters 
    of the waveform in real time, standard configuration includes the navigation knob
    Pacemaker detection, electrosurgical interference proof
    Defibrillation protection and defibrillation synchronization
    SD Card slot enable memory extension for large data storage
    USB and serial port enable abundant future upgrades
    VGA output and Analog output
    OxyCRG available to judge the respiration and circulation function for neonate
    Large font display
    Arrhythmia and S-T segment analysis
    Built-in rechargeable Li-ion (6hours maximum with 2x 4200mAh batteries)
    Nurse Call function and bi-directional communication with MFM-CMS central station
    Multi language: English/Simplified Chinese/Traditional Chinese/Spanish/Czech/Italian/Polish/French/German/Brazilian/Portugese/Russian.
    Standard configuration: 3/5-lead ECG, NIBP, SpO2, RESP, 2-Temp, PR
    Optional: 12-lead ECG, Nellcor-SpO2, Omron NIBP, 2-IBP/4-IBP, Mainstream/Side stream CO2, 
    CO, Multi-Gas, Full Touch Screen, Thermal Recorder, WLAN Accessory
    Large Data Storage: 120 hours graphic and tabular trends of all parameters; 1200 sets of NIBP
     measurement and 70 alarm events; 2-USB ports and SC card slot ensure data storage/function upgrades.
    Clinical data can be viewed and managed via Patientcare viewer software (To be released)
    Clinical network: bi-directional communication with CMS-2000 central station by wired or wireless connection
    Complete and flexible mounting solution for ICU and operation room. High-end parameters for more flexible clinical applications.


    Patient Monitor Support:
    DC power input and to adapt to a variety of environmental conditions
    External recorder can print the recorded waveform data and other information in real time
    External the EtCO2 interface can be connected to the mainstream CO2 module to facilitate the clinical needs
    Drug concentration in titration calculations to support the event functions to customize and record
    Up to 32 TCP / IP network connection composed of a central information system
    Wireless networking
    15 inch touch screen patient monitor,Slim patient monitor ,Meditech patient monitor

    Options 
    Invasive Blood Pressure (IBP)
    Carbon dioxide (CO2) Module
    Masimo SpO2 Module
    Thermal Printer

    Built-in Wifi (MD9015T)

Product Specifications

    Electrocardiogram (ECG)
    Input mode Five-lead ECG input / three-lead ECG input
    Lead selection I, II, III, aVR, aVL, aVF, V-
    Gain selection AUTO,0.25X,0.5X,1X,2X,4X
    Operating mode Diagnostic mode, monitoring mode, operation mode
    Calibration signal 1mV ±5%
    Protection The isolation voltage 4000V, 50/60Hz. Anti-high-frequency interference of electric knife. Anti-defibrillation protection.
    Pacemaker at the same time use Support
    Electrode off instructions Each electrode (RL excluded)
    Scanning speed 12.5mm/s, 25 mm/s, 50 mm/s
    Common mode rejection ratio ≥90dB
    Input impedance ≥5.0MΩ
     
    Heartrate (HR)                          

    Measuring range Adult 30 ~ 300bpm, newborns 30 to 350 bpm
    Resolution 1 bpm
    Accuracy < ±5bpm
    Heart rate sensitivity ≥0.20mVp-p
    Alarm range 15 to 350 bpm, upper and lower limits
    Alarm indication Audible and visual alarm, display the data and parameters flashing display alarm 
    message text prompts, there are three levels of alarm
     
    Pulse Oximetry
    SPO2 Range:                          1% to 99%
    Resolution:                              1%
    Accuracy:                                70% to 99% range: ±2%;
    1% to 69% range:                    ±5%;
    Method:                                  Dual wavelength LED 

    ST-segment

    ST-segment measurement range -2.0~+2.0mV
    ST-segment measurement accuracy Within the range of -0.8 ~ +0.8 mV: ± 0.02mV or reading± 10%,
    Whichever is greater; Other range: undefined
    Resolution 0.01mV
    ST-segment update cycle 10s
    ST segment alarm range -2.0 ~ +2.0 MV, the upper and lower limits continuously adjustable
     
    Arrhythmia analysis

    Arrhythmia types 13 kinds of (scalable)
     
    Respiration Rate
    Rate:                                  2 to 120 bpm (CO2)
    Resolution:                        1 breath/min
     
    NIBP (Non-invasive Blood Pressure)
    Technique:                        Oscillometric during inflation
    Range:                               Adult/Pediatric: 40 to 240 mmHg
    Neonate:                           10 to 150 mmHg
    Measurement Cycle:         < 40 sec. typical
    Automatic Measurement
    Cycles (Selectable):          2, 3, 5, 10, 15, 30 min; 1, 2, 4 hr
    STAT Mode:                      5 minutes of continuous readings
    Max. Allowable Cuff
    Pressure: Adult:                300mmHg
    Pediatric:                          240 mmHg;
    Neonate:                          150 mmHg
    Resolution:                       1 mmHg
    Transducer Accuracy:       ±2 mmHg or ±2% of reading,
    Heart (Pulse) Rate
    Range:                              30 to 250 bpm
    Accuracy:                        ±2 bpm or ±3%, whichever is greater


    EtCO2 technological index


    Interface form External interface
    Range of
    measurement
    0~10 kPa or (0~75) mmHg
    precision(error) (0kPa~5.4kPa) the error is ±0.3kPa、
    (5.5kPa~10.0kPa)the error is ±0.7kPa
    Alarm upper limit 2.1~13.2 kPa
    Alarm lower limit 0~6.5 kPa
    time 12s
    EtCO2 exist
    Unit kPa / mmHg / %
    Compatibility  
    external main stream and side stream are compatible

    IBP technological index
     
    Measured channel Single way and double way are optional
    Range of
    measurement
    SYS: 6.7kPa~32.0kPa(50mmHg~240mmHg)
    MEAN: 3.4kPa~26.6kPa(26mmHg~200mmHg)
    DIA: 2.9kPa~24.0kPa(22mmHg~180mmHg)
    Mean error ≤±0.4kPa (±3mmHg)
    Range of alarm SYS:  upper limit   0.3kPa~53.3kPa(2mmHg~400mmHg)
    Lower limit   0kPa~53.1kPa(0mmHg~398mmHg)
    MEAN:  upper limit    0.3kPa~53.3kPa (2mmHg~400mmHg)
    Lower limit   0kPa~53.1kPa (0mmHg~398mmHg)
    DIA:    upper limit    0.3kPa~53.3kPa ( 2mmHg~400mmHg)
              Lower limit   0kPa~53.1kPa ( 0mmHg~398mmHg)
    Measured unit mmHg/kPa
    Specification of
    sensor
    DPT-248(space)
    Consumables mode SCW-UTAH-01
       
       
       
       
       

Related Products

Promoted Products

Meditech Equipment Co.,Ltd is part of Meditech Group. Product(s) described may not be licensed or available for sale in all countries. Sonotech, Sonovet, iSonic, FOs2pro, Dolphi, Defi, HeartRec,miniScan,Cardios,SpirOx,iBreath, Meditech and all corresponding design marks are trademarks of Meditech. The symbol indicates the trademark is registered. Patent and Trademark Office and certain other countries. All other names and marks mentioned are the trade names, trademarks or service marks of their respective owners. Please see the Instructions for Use for a complete listing of the indications, contraindications, warnings and precautions.

Legal notice Terms and conditions Cookie policy Privacy Policy Professional organisations Careers

2024© Copyright احدى فروع شركة ميديتك جروب و المركز الرئيسي الإقليمي لمنطقتي الشرق الأوسط و شمال إفريقيا | Meditech Group (Middle East branch).

Meditech Group

To Top