MD06x Advanced blood pressure with Oximetry,
3 users data record and alarm setting,
Meditech 06X have a large color screen, and come with PC software to Show trend graph,
analytical diagrams Edit add notes to every piece of data by connecting the device with PC through USB port.


Product Features

    2.8" TFT color screen 320x240 pixels
    NIBP measurement
    SpO2 measurement
    Trusted accuracy clinically validated
    Oscillometry with step deflation
    Multi-user mode (Adult Child New born)
    Built-in flash memory up to 100 items for each user (3 User)
    Built-in BP Alarms
    Automatic cuff inflation and deflation
    Reliable power supply with or AA battery
    Battery status indicator
    Data Review: List and Trend
    Setting for Time .Date .mHg/kPa .Alarm .User type .Language
    Sync with PC-based Software via USB port accurate analysis systems
     



     

     
    Comes with PC Software for data transfer
    Show trend graph analytical diagrams Edit add notes to every piece of data 
    Edit examinees information medical advices condition descriptions present drug usage info etc. 
    Print report support print preview
     
    06X Soft


Product Specifications

    Measurement Accuracy
    SpO2 70%~100%    ±2%
    0%~69%      undefined
    PR ±2bpm or ±2% (select larger) 
    Measurement Performance In Weak Filling Condition
    pulse-filling ratio : 0.4% SpO2 error ±4%
    pulse rate error ±2bpm or ±2% (select larger)
    Optical Sensor
    Red light wavelength is 660nm 6.65mW
    Infrared wavelength is 880nm 6.75mW  
    Display 2.8'' color LCD Display
     
    NIBP Specifications
    Measurement Method Oscillometric method

    Working modes Automatic
    Measurement Range Pressure adult 0~290mmHg(0~38.6kPa)
    pediatric 0~235mmHg(0~31.3kPa)
    neonatal 0~140mmHg(0~18.6kPa)
    Pulse: 40~240/min
    Inflation adult 160mmHg
    pediatric 120mmHg
    neonatal 70mmHg
    Alarm Range adult mode SYS ALM: 40~270 mmHg 
    DIA ALM: 10~215 mmHg
    pediatric mode SYS ALM: 40~200 mmHg
    DIA ALM: 10~150 mmHg
    neonatal mode SYS ALM: 40~135 mmHg
    DIA ALM: 10~100 mmHg
    Overpressure protect adult mode 295±5mmHg
    pediatric mode 240±5mmHg
    neonatal mode 145±5mmHg

    Resolution
    Pressure 1mmHg

    Measurement Accuracy
    Cuff Pressure Accuracy ±3mmHg
    Error The BP Value of the device is equivalence with the measurement value of Stethoscopy. The error meets all the conditions in the YY0667-2008.
    Operating Temperature/ Humidity +5oC~40 oC      .  15%RH~80%RH
    Transport and Storage Temperature/Humidity -20oC~+55oC     .  ≤95%RH
    Atmospheric pressure 80kPa~105kPa
    Battery 4 "AA" alkaline batteries AC Adapter separately sold
    Main Unit Dimensions 130(L)*110(W)*80mm(H)  
    Main Unit Weight 300gram
     

Related Products

Promoted Products

Meditech Equipment Co.,Ltd is part of Meditech Group. Product(s) described may not be licensed or available for sale in all countries. Sonotech, Sonovet, iSonic, FOs2pro, Dolphi, Defi, HeartRec,miniScan,Cardios,SpirOx,iBreath, Meditech and all corresponding design marks are trademarks of Meditech. The symbol indicates the trademark is registered. Patent and Trademark Office and certain other countries. All other names and marks mentioned are the trade names, trademarks or service marks of their respective owners. Please see the Instructions for Use for a complete listing of the indications, contraindications, warnings and precautions.