ECG 
	Lead Mode 5 Leads ( R, L, F, N, C or RA, LA, LL, RL, V)
	Lead selection I, II, III, avR, avL, avF, V, 
	Waveform 2 ch 
	Lead mode 3 Leads ( R, L, F or RA, LA, LL)
	Lead selection I, II, III, 
	Waveform 1 ch 
	Gain x2.5mm/mV, x5.0mm/mV, x10mm/mV, x20mm/mV, auto 
	HR and Alarm  
	      Range 
	      Adult 15 ~ 300 bpm
	      Neo/Ped 15 ~ 350 bpm
	      Accuracy ± 1% or ± 1bpm, which great
	      Resolution 1 bpm
	     Sensitivity > 200 (uV ) 
	Differential Input Impedance > 5 MΩ
	CMRR
	      Monitor > 105 dB 
	     Operation > 105 dB
	     Diagnosis > 85 dB
	Electrode offset potential ±300mV
	Leakage Current < 10 uA
	Baseline Recovery < 3 S After Defi.
	ECG Signal Range ±8 m V ( Vp-p ) 
	Bandwidth    
	     Surgery 1 ~ 15 Hz
	     Monitor 0.5 ~ 35 Hz
	     Diagnostic 0.05 ~ 100 Hz
	Calibration Signal 1 (mV ), Accuracy : ±5%  
	ST Segment Monitoring Range   
	Measure and Alarm -2.0 ~ +2.0 mV
	ARR Detecting  
	Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC 
	
		    Alarm Available
	
		    Review Available 
	
		
		RESPARATION 
	
		Method Impedance between RF (RA-LL)
	
		Differential Input Impedance >2.5 MΩ
	
		Measuring Impedance Range: 0.3~5.0Ω
	
		Base line Impedance Range: 0 – 2.5 KΩ
	
		Bandwidth 0.3 ~ 2.5 Hz
	
		Resp. Rate 
	
		Measuring and Alarm Range 
	
		     Adult 0 ~ 120 rpm
	
		     Neo/Ped 0 ~ 150 rpm 
	
		Resolution 1 rpm
	
		Accuracy ±2 rpm 
	
		Apean Alarm 10 ~ 40 S 
	
		
		NIBP 
	
		Method Oscillometric
	
		Mode Manual, Auto, STAT
	
		Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min)
	
		Measuring Period in STAT Mode 5 Min
	
		Pulse Rate Range 40 ~ 240 bpm
	
		Alarm Type SYS, DIA, MEAN
	
		Measuring and alarm range    
	
		Adult Mode
	
		             SYS 40 ~ 270 mmHg
	
		             DIA 10 ~ 215 mmHg
	
		             MEAN 20 ~ 235 mmHg 
	
		Pediatric Mode
	
		             SYS 40 ~ 200 mmHg
	
		             DIA 10 ~ 150 mmHg 
	
		             MEAN 20 ~ 165 mmHg
		Neonatal Mode 
	
		             SYS 40 ~ 135 mmHg
	
		             DIA 10 ~ 100 mmHg
	
		             MEAN 20 ~ 110 mmHg 
	
		Resolution 
	
		              Pressure 1mmHg
	
		              Accuracy
	
		              Pressure  
	
		Maximum Mean error ±5mmHg
	
		Maximum Standard deviation ±8mmHg 
	
		Overpressure Protection
	
		             Adult Mode 297±3 mmHg 
	
		             Pediatric Mode 240±3 mmHg
	
		             Neonatal Mode 147±3 mmHg 
	
		
		SpO2 
	
		Measuring Range 0 ~ 100 %
	
		Alarm Range 0 ~ 100 %
	
		Resolution 1 %
	
		Accuracy 70% ~ 100% ±2 %
	
		                                     0% ~ 69% unspecified
	
		Actualization interval about 1 Sec.
	
		Alarm Delay 10 Sec. 
	
		Pulse Rate 
	
		Measuring and Alarm Range 0~254bpm
	
		Resolution 1bpm 
	
		 Accuracy ±2bpm 
	
		
		TEMPERATURE 
	
		Channel 1
	
		Measuring and Alarm Range 0 ~ 50 C
	
		Resolution 0.1~C
	
		Accuracy ±0.1~C 
	
		Actualization interval about 1 Sec.
	
		Average Time Constant < 10 Sec. 
	
		
		IBP 
	
		Label ART, PA, CVP, RAP, LAP, ICP, P1, P2
	
		Measuring and alarm range 
	
		    ART 0 ~ 300 mmHg
	
		    PA -6 ~ 120 mmHg
	
		    CVP/RAP/LAP/ICP -10 ~ 40 mmHg
	
		    P1/P2 -10 ~ 300 mmHg     
	
		Press Sensor 
	
		Sensitivity 5 uV/V/mmHg 
		
			   Impedance 300-3000Ω
		
			   Resolution 1 mmHg 
		
			   Accuracy ±2% or ±1mmHg, which great about 1 Sec. 
		
			 
	 
 
	
		ECG 
	
		Lead Mode 5 Leads ( R, L, F, N, C or RA, LA, LL, RL, V)
	
		Lead selection I, II, III, avR, avL, avF, V, 
	
		Waveform 2 ch 
	
		Lead mode 3 Leads ( R, L, F or RA, LA, LL)
	
		Lead selection I, II, III, 
	
		Waveform 1 ch 
	
		Gain x2.5mm/mV, x5.0mm/mV, x10mm/mV, x20mm/mV, auto 
	
		HR and Alarm  
	
		      Range 
	
		      Adult 15 ~ 300 bpm
	
		      Neo/Ped 15 ~ 350 bpm
	
		      Accuracy ± 1% or ± 1bpm, which great
	
		      Resolution 1 bpm
	
		     Sensitivity > 200 (uV ) 
	
		Differential Input Impedance > 5 MΩ
	
		CMRR
	
		      Monitor > 105 dB 
	
		     Operation > 105 dB
	
		     Diagnosis > 85 dB
	
		Electrode offset potential ±300mV
	
		Leakage Current < 10 uA
	
		Baseline Recovery < 3 S After Defi.
	
		ECG Signal Range ±8 m V ( Vp-p ) 
	
		Bandwidth    
	
		     Surgery 1 ~ 15 Hz
	
		     Monitor 0.5 ~ 35 Hz
	
		     Diagnostic 0.05 ~ 100 Hz
	
		Calibration Signal 1 (mV ), Accuracy : ±5%  
	
		ST Segment Monitoring Range   
	
		Measure and Alarm -2.0 ~ +2.0 mV
	
		ARR Detecting  
	
		Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC 
	
		
			    Alarm Available
		
			    Review Available 
		
			
			RESPARATION 
		
			Method Impedance between RF (RA-LL)
		
			Differential Input Impedance >2.5 MΩ
		
			Measuring Impedance Range: 0.3~5.0Ω
		
			Base line Impedance Range: 0 – 2.5 KΩ
		
			Bandwidth 0.3 ~ 2.5 Hz
		
			Resp. Rate 
		
			Measuring and Alarm Range 
		
			     Adult 0 ~ 120 rpm
		
			     Neo/Ped 0 ~ 150 rpm 
		
			Resolution 1 rpm
		
			Accuracy ±2 rpm 
		
			Apean Alarm 10 ~ 40 S 
		
			
			NIBP 
		
			Method Oscillometric
		
			Mode Manual, Auto, STAT
		
			Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min)
		
			Measuring Period in STAT Mode 5 Min
		
			Pulse Rate Range 40 ~ 240 bpm
		
			Alarm Type SYS, DIA, MEAN
		
			Measuring and alarm range    
		
			Adult Mode
		
			             SYS 40 ~ 270 mmHg
		
			             DIA 10 ~ 215 mmHg
		
			             MEAN 20 ~ 235 mmHg 
		
			Pediatric Mode
		
			             SYS 40 ~ 200 mmHg
		
			             DIA 10 ~ 150 mmHg 
		
			             MEAN 20 ~ 165 mmHg
			Neonatal Mode 
		
			             SYS 40 ~ 135 mmHg
		
			             DIA 10 ~ 100 mmHg
		
			             MEAN 20 ~ 110 mmHg 
		
			Resolution 
		
			              Pressure 1mmHg
		
			              Accuracy
		
			              Pressure  
		
			Maximum Mean error ±5mmHg
		
			Maximum Standard deviation ±8mmHg 
		
			Overpressure Protection
		
			             Adult Mode 297±3 mmHg 
		
			             Pediatric Mode 240±3 mmHg
		
			             Neonatal Mode 147±3 mmHg 
		
			
			SpO2 
		
			Measuring Range 0 ~ 100 %
		
			Alarm Range 0 ~ 100 %
		
			Resolution 1 %
		
			Accuracy 70% ~ 100% ±2 %
		
			                                     0% ~ 69% unspecified
		
			Actualization interval about 1 Sec.
		
			Alarm Delay 10 Sec. 
		
			Pulse Rate 
		
			Measuring and Alarm Range 0~254bpm
		
			Resolution 1bpm 
		
			 Accuracy ±2bpm 
		
			
			TEMPERATURE 
		
			Channel 1
		
			Measuring and Alarm Range 0 ~ 50 C
		
			Resolution 0.1~C
		
			Accuracy ±0.1~C 
		
			Actualization interval about 1 Sec.
		
			Average Time Constant < 10 Sec. 
		
			
			IBP 
		
			Label ART, PA, CVP, RAP, LAP, ICP, P1, P2
		
			Measuring and alarm range 
		
			    ART 0 ~ 300 mmHg
		
			    PA -6 ~ 120 mmHg
		
			    CVP/RAP/LAP/ICP -10 ~ 40 mmHg
		
			    P1/P2 -10 ~ 300 mmHg     
		
			Press Sensor 
		
			Sensitivity 5 uV/V/mmHg 
			
				   Impedance 300-3000Ω
			
				   Resolution 1 mmHg 
			
				   Accuracy ±2% or ±1mmHg, which great about 1 Sec.