MD908D
15 inch color TFT display Multi parameters Patient Monitor;      
A variety of display interfaces: large font, standard, full lead, respiratory oxygen, blood pressure trend;                        
Built-in rechargeable battery, work for 2 hours(more);   

15 inch color TFT display Multi parameters Patient Monitor;      
A variety of display interfaces: large font, standard, full lead, respiratory oxygen, blood pressure trend;                        
Built-in rechargeable battery, work for 2 hours(more);   

Read More
High resolution 15” color TFT display
12 Channel waves.9 Parameters : ECG, Resp, NIBP, SpO2, TEMP,
Light weight, compact and portable
Built-in rechargeable lithium battery 
Unique human voice alarm
Rich screen layouts: Standard, Big font, Trend table, NIBP trend, ECG full lead, OxyCRG, IoC Layout
Powerful storage: Maximum 720 Hour graphic and tabular trends of all
Parameters, up to 10000 NIBP review ,200alarm review ,2Hours of all waveforms
System software can be upgraded through network
Data and waveforms color be adjustable
Arrhythmia analysis and ST segment analysis
Suitable for adult, paediatric and neonatal patient
ECG 
Lead Mode 5 Leads ( R, L, F, N, C or RA, LA, LL, RL, V)
Lead selection I, II, III, avR, avL, avF, V, 
Waveform 2 ch 
Lead mode 3 Leads ( R, L, F or RA, LA, LL)
Lead selection I, II, III, 
Waveform 1 ch 
Gain x2.5mm/mV, x5.0mm/mV, x10mm/mV, x20mm/mV, auto 
HR and Alarm  
      Range 
      Adult 15 ~ 300 bpm
      Neo/Ped 15 ~ 350 bpm
      Accuracy ± 1% or ± 1bpm, which great
      Resolution 1 bpm
     Sensitivity > 200 (uV ) 
Differential Input Impedance > 5 MΩ
CMRR
      Monitor > 105 dB 
     Operation > 105 dB
     Diagnosis > 85 dB
Electrode offset potential ±300mV
Leakage Current < 10 uA
Baseline Recovery < 3 S After Defi.
ECG Signal Range ±8 m V ( Vp-p ) 
Bandwidth    
     Surgery 1 ~ 15 Hz
     Monitor 0.5 ~ 35 Hz
     Diagnostic 0.05 ~ 100 Hz
Calibration Signal 1 (mV ), Accuracy : ±5%  
ST Segment Monitoring Range   
Measure and Alarm -2.0 ~ +2.0 mV
ARR Detecting  
Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC 
    Alarm Available
    Review Available 

RESPARATION 
Method Impedance between RF (RA-LL)
Differential Input Impedance >2.5 MΩ
Measuring Impedance Range: 0.3~5.0Ω
Base line Impedance Range: 0 – 2.5 KΩ
Bandwidth 0.3 ~ 2.5 Hz
Resp. Rate 
Measuring and Alarm Range 
     Adult 0 ~ 120 rpm
     Neo/Ped 0 ~ 150 rpm 
Resolution 1 rpm
Accuracy ±2 rpm 
Apean Alarm 10 ~ 40 S 

NIBP 
Method Oscillometric
Mode Manual, Auto, STAT
Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min)
Measuring Period in STAT Mode 5 Min
Pulse Rate Range 40 ~ 240 bpm
Alarm Type SYS, DIA, MEAN
Measuring and alarm range    
Adult Mode
             SYS 40 ~ 270 mmHg
             DIA 10 ~ 215 mmHg
             MEAN 20 ~ 235 mmHg 
Pediatric Mode
             SYS 40 ~ 200 mmHg
             DIA 10 ~ 150 mmHg 
             MEAN 20 ~ 165 mmHg
Neonatal Mode 
             SYS 40 ~ 135 mmHg
             DIA 10 ~ 100 mmHg
             MEAN 20 ~ 110 mmHg 
Resolution 
              Pressure 1mmHg
              Accuracy
              Pressure  
Maximum Mean error ±5mmHg
Maximum Standard deviation ±8mmHg 
Overpressure Protection
             Adult Mode 297±3 mmHg 
             Pediatric Mode 240±3 mmHg
             Neonatal Mode 147±3 mmHg 

SpO2 
Measuring Range 0 ~ 100 %
Alarm Range 0 ~ 100 %
Resolution 1 %
Accuracy 70% ~ 100% ±2 %
                                     0% ~ 69% unspecified
Actualization interval about 1 Sec.
Alarm Delay 10 Sec. 
Pulse Rate 
Measuring and Alarm Range 0~254bpm
Resolution 1bpm 
 Accuracy ±2bpm 

TEMPERATURE 
Channel 1
Measuring and Alarm Range 0 ~ 50 C
Resolution 0.1~C
Accuracy ±0.1~C 
Actualization interval about 1 Sec.
Average Time Constant < 10 Sec. 

IBP 
Label ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range 
    ART 0 ~ 300 mmHg
    PA -6 ~ 120 mmHg
    CVP/RAP/LAP/ICP -10 ~ 40 mmHg
    P1/P2 -10 ~ 300 mmHg     
Press Sensor 
Sensitivity 5 uV/V/mmHg 
   Impedance 300-3000Ω
   Resolution 1 mmHg 
   Accuracy ±2% or ±1mmHg, which great about 1 Sec. 
 

ECG 
Lead Mode 5 Leads ( R, L, F, N, C or RA, LA, LL, RL, V)
Lead selection I, II, III, avR, avL, avF, V, 
Waveform 2 ch 
Lead mode 3 Leads ( R, L, F or RA, LA, LL)
Lead selection I, II, III, 
Waveform 1 ch 
Gain x2.5mm/mV, x5.0mm/mV, x10mm/mV, x20mm/mV, auto 
HR and Alarm  
      Range 
      Adult 15 ~ 300 bpm
      Neo/Ped 15 ~ 350 bpm
      Accuracy ± 1% or ± 1bpm, which great
      Resolution 1 bpm
     Sensitivity > 200 (uV ) 
Differential Input Impedance > 5 MΩ
CMRR
      Monitor > 105 dB 
     Operation > 105 dB
     Diagnosis > 85 dB
Electrode offset potential ±300mV
Leakage Current < 10 uA
Baseline Recovery < 3 S After Defi.
ECG Signal Range ±8 m V ( Vp-p ) 
Bandwidth    
     Surgery 1 ~ 15 Hz
     Monitor 0.5 ~ 35 Hz
     Diagnostic 0.05 ~ 100 Hz
Calibration Signal 1 (mV ), Accuracy : ±5%  
ST Segment Monitoring Range   
Measure and Alarm -2.0 ~ +2.0 mV
ARR Detecting  
Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC 
    Alarm Available
    Review Available 

RESPARATION 
Method Impedance between RF (RA-LL)
Differential Input Impedance >2.5 MΩ
Measuring Impedance Range: 0.3~5.0Ω
Base line Impedance Range: 0 – 2.5 KΩ
Bandwidth 0.3 ~ 2.5 Hz
Resp. Rate 
Measuring and Alarm Range 
     Adult 0 ~ 120 rpm
     Neo/Ped 0 ~ 150 rpm 
Resolution 1 rpm
Accuracy ±2 rpm 
Apean Alarm 10 ~ 40 S 

NIBP 
Method Oscillometric
Mode Manual, Auto, STAT
Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min)
Measuring Period in STAT Mode 5 Min
Pulse Rate Range 40 ~ 240 bpm
Alarm Type SYS, DIA, MEAN
Measuring and alarm range    
Adult Mode
             SYS 40 ~ 270 mmHg
             DIA 10 ~ 215 mmHg
             MEAN 20 ~ 235 mmHg 
Pediatric Mode
             SYS 40 ~ 200 mmHg
             DIA 10 ~ 150 mmHg 
             MEAN 20 ~ 165 mmHg
Neonatal Mode 
             SYS 40 ~ 135 mmHg
             DIA 10 ~ 100 mmHg
             MEAN 20 ~ 110 mmHg 
Resolution 
              Pressure 1mmHg
              Accuracy
              Pressure  
Maximum Mean error ±5mmHg
Maximum Standard deviation ±8mmHg 
Overpressure Protection
             Adult Mode 297±3 mmHg 
             Pediatric Mode 240±3 mmHg
             Neonatal Mode 147±3 mmHg 

SpO2 
Measuring Range 0 ~ 100 %
Alarm Range 0 ~ 100 %
Resolution 1 %
Accuracy 70% ~ 100% ±2 %
                                     0% ~ 69% unspecified
Actualization interval about 1 Sec.
Alarm Delay 10 Sec. 
Pulse Rate 
Measuring and Alarm Range 0~254bpm
Resolution 1bpm 
 Accuracy ±2bpm 

TEMPERATURE 
Channel 1
Measuring and Alarm Range 0 ~ 50 C
Resolution 0.1~C
Accuracy ±0.1~C 
Actualization interval about 1 Sec.
Average Time Constant < 10 Sec. 

IBP 
Label ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range 
    ART 0 ~ 300 mmHg
    PA -6 ~ 120 mmHg
    CVP/RAP/LAP/ICP -10 ~ 40 mmHg
    P1/P2 -10 ~ 300 mmHg     
Press Sensor 
Sensitivity 5 uV/V/mmHg 
   Impedance 300-3000Ω
   Resolution 1 mmHg 
   Accuracy ±2% or ±1mmHg, which great about 1 Sec. 

Product Features

    High resolution 15” color TFT display
    12 Channel waves.9 Parameters : ECG, Resp, NIBP, SpO2, TEMP,
    Light weight, compact and portable
    Built-in rechargeable lithium battery 
    Unique human voice alarm
    Rich screen layouts: Standard, Big font, Trend table, NIBP trend, ECG full lead, OxyCRG, IoC Layout
    Powerful storage: Maximum 720 Hour graphic and tabular trends of all
    Parameters, up to 10000 NIBP review ,200alarm review ,2Hours of all waveforms
    System software can be upgraded through network
    Data and waveforms color be adjustable
    Arrhythmia analysis and ST segment analysis
    Suitable for adult, paediatric and neonatal patient

Product Specifications

    ECG 
    Lead Mode 5 Leads ( R, L, F, N, C or RA, LA, LL, RL, V)
    Lead selection I, II, III, avR, avL, avF, V, 
    Waveform 2 ch 
    Lead mode 3 Leads ( R, L, F or RA, LA, LL)
    Lead selection I, II, III, 
    Waveform 1 ch 
    Gain x2.5mm/mV, x5.0mm/mV, x10mm/mV, x20mm/mV, auto 
    HR and Alarm  
          Range 
          Adult 15 ~ 300 bpm
          Neo/Ped 15 ~ 350 bpm
          Accuracy ± 1% or ± 1bpm, which great
          Resolution 1 bpm
         Sensitivity > 200 (uV ) 
    Differential Input Impedance > 5 MΩ
    CMRR
          Monitor > 105 dB 
         Operation > 105 dB
         Diagnosis > 85 dB
    Electrode offset potential ±300mV
    Leakage Current < 10 uA
    Baseline Recovery < 3 S After Defi.
    ECG Signal Range ±8 m V ( Vp-p ) 
    Bandwidth    
         Surgery 1 ~ 15 Hz
         Monitor 0.5 ~ 35 Hz
         Diagnostic 0.05 ~ 100 Hz
    Calibration Signal 1 (mV ), Accuracy : ±5%  
    ST Segment Monitoring Range   
    Measure and Alarm -2.0 ~ +2.0 mV
    ARR Detecting  
    Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC 
        Alarm Available
        Review Available 

    RESPARATION 
    Method Impedance between RF (RA-LL)
    Differential Input Impedance >2.5 MΩ
    Measuring Impedance Range: 0.3~5.0Ω
    Base line Impedance Range: 0 – 2.5 KΩ
    Bandwidth 0.3 ~ 2.5 Hz
    Resp. Rate 
    Measuring and Alarm Range 
         Adult 0 ~ 120 rpm
         Neo/Ped 0 ~ 150 rpm 
    Resolution 1 rpm
    Accuracy ±2 rpm 
    Apean Alarm 10 ~ 40 S 

    NIBP 
    Method Oscillometric
    Mode Manual, Auto, STAT
    Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min)
    Measuring Period in STAT Mode 5 Min
    Pulse Rate Range 40 ~ 240 bpm
    Alarm Type SYS, DIA, MEAN
    Measuring and alarm range    
    Adult Mode
                 SYS 40 ~ 270 mmHg
                 DIA 10 ~ 215 mmHg
                 MEAN 20 ~ 235 mmHg 
    Pediatric Mode
                 SYS 40 ~ 200 mmHg
                 DIA 10 ~ 150 mmHg 
                 MEAN 20 ~ 165 mmHg
    Neonatal Mode 
                 SYS 40 ~ 135 mmHg
                 DIA 10 ~ 100 mmHg
                 MEAN 20 ~ 110 mmHg 
    Resolution 
                  Pressure 1mmHg
                  Accuracy
                  Pressure  
    Maximum Mean error ±5mmHg
    Maximum Standard deviation ±8mmHg 
    Overpressure Protection
                 Adult Mode 297±3 mmHg 
                 Pediatric Mode 240±3 mmHg
                 Neonatal Mode 147±3 mmHg 

    SpO2 
    Measuring Range 0 ~ 100 %
    Alarm Range 0 ~ 100 %
    Resolution 1 %
    Accuracy 70% ~ 100% ±2 %
                                         0% ~ 69% unspecified
    Actualization interval about 1 Sec.
    Alarm Delay 10 Sec. 
    Pulse Rate 
    Measuring and Alarm Range 0~254bpm
    Resolution 1bpm 
     Accuracy ±2bpm 

    TEMPERATURE 
    Channel 1
    Measuring and Alarm Range 0 ~ 50 C
    Resolution 0.1~C
    Accuracy ±0.1~C 
    Actualization interval about 1 Sec.
    Average Time Constant < 10 Sec. 

    IBP 
    Label ART, PA, CVP, RAP, LAP, ICP, P1, P2
    Measuring and alarm range 
        ART 0 ~ 300 mmHg
        PA -6 ~ 120 mmHg
        CVP/RAP/LAP/ICP -10 ~ 40 mmHg
        P1/P2 -10 ~ 300 mmHg     
    Press Sensor 
    Sensitivity 5 uV/V/mmHg 
       Impedance 300-3000Ω
       Resolution 1 mmHg 
       Accuracy ±2% or ±1mmHg, which great about 1 Sec. 
     

    ECG 
    Lead Mode 5 Leads ( R, L, F, N, C or RA, LA, LL, RL, V)
    Lead selection I, II, III, avR, avL, avF, V, 
    Waveform 2 ch 
    Lead mode 3 Leads ( R, L, F or RA, LA, LL)
    Lead selection I, II, III, 
    Waveform 1 ch 
    Gain x2.5mm/mV, x5.0mm/mV, x10mm/mV, x20mm/mV, auto 
    HR and Alarm  
          Range 
          Adult 15 ~ 300 bpm
          Neo/Ped 15 ~ 350 bpm
          Accuracy ± 1% or ± 1bpm, which great
          Resolution 1 bpm
         Sensitivity > 200 (uV ) 
    Differential Input Impedance > 5 MΩ
    CMRR
          Monitor > 105 dB 
         Operation > 105 dB
         Diagnosis > 85 dB
    Electrode offset potential ±300mV
    Leakage Current < 10 uA
    Baseline Recovery < 3 S After Defi.
    ECG Signal Range ±8 m V ( Vp-p ) 
    Bandwidth    
         Surgery 1 ~ 15 Hz
         Monitor 0.5 ~ 35 Hz
         Diagnostic 0.05 ~ 100 Hz
    Calibration Signal 1 (mV ), Accuracy : ±5%  
    ST Segment Monitoring Range   
    Measure and Alarm -2.0 ~ +2.0 mV
    ARR Detecting  
    Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC 
        Alarm Available
        Review Available 

    RESPARATION 
    Method Impedance between RF (RA-LL)
    Differential Input Impedance >2.5 MΩ
    Measuring Impedance Range: 0.3~5.0Ω
    Base line Impedance Range: 0 – 2.5 KΩ
    Bandwidth 0.3 ~ 2.5 Hz
    Resp. Rate 
    Measuring and Alarm Range 
         Adult 0 ~ 120 rpm
         Neo/Ped 0 ~ 150 rpm 
    Resolution 1 rpm
    Accuracy ±2 rpm 
    Apean Alarm 10 ~ 40 S 

    NIBP 
    Method Oscillometric
    Mode Manual, Auto, STAT
    Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min)
    Measuring Period in STAT Mode 5 Min
    Pulse Rate Range 40 ~ 240 bpm
    Alarm Type SYS, DIA, MEAN
    Measuring and alarm range    
    Adult Mode
                 SYS 40 ~ 270 mmHg
                 DIA 10 ~ 215 mmHg
                 MEAN 20 ~ 235 mmHg 
    Pediatric Mode
                 SYS 40 ~ 200 mmHg
                 DIA 10 ~ 150 mmHg 
                 MEAN 20 ~ 165 mmHg
    Neonatal Mode 
                 SYS 40 ~ 135 mmHg
                 DIA 10 ~ 100 mmHg
                 MEAN 20 ~ 110 mmHg 
    Resolution 
                  Pressure 1mmHg
                  Accuracy
                  Pressure  
    Maximum Mean error ±5mmHg
    Maximum Standard deviation ±8mmHg 
    Overpressure Protection
                 Adult Mode 297±3 mmHg 
                 Pediatric Mode 240±3 mmHg
                 Neonatal Mode 147±3 mmHg 

    SpO2 
    Measuring Range 0 ~ 100 %
    Alarm Range 0 ~ 100 %
    Resolution 1 %
    Accuracy 70% ~ 100% ±2 %
                                         0% ~ 69% unspecified
    Actualization interval about 1 Sec.
    Alarm Delay 10 Sec. 
    Pulse Rate 
    Measuring and Alarm Range 0~254bpm
    Resolution 1bpm 
     Accuracy ±2bpm 

    TEMPERATURE 
    Channel 1
    Measuring and Alarm Range 0 ~ 50 C
    Resolution 0.1~C
    Accuracy ±0.1~C 
    Actualization interval about 1 Sec.
    Average Time Constant < 10 Sec. 

    IBP 
    Label ART, PA, CVP, RAP, LAP, ICP, P1, P2
    Measuring and alarm range 
        ART 0 ~ 300 mmHg
        PA -6 ~ 120 mmHg
        CVP/RAP/LAP/ICP -10 ~ 40 mmHg
        P1/P2 -10 ~ 300 mmHg     
    Press Sensor 
    Sensitivity 5 uV/V/mmHg 
       Impedance 300-3000Ω
       Resolution 1 mmHg 
       Accuracy ±2% or ±1mmHg, which great about 1 Sec. 