15 inch color TFT display Multi parameters Patient Monitor;      
A variety of display interfaces: large font, standard, full lead, respiratory oxygen, blood pressure trend;                        
Built-in rechargeable battery, work for 2 hours(more);   
 

Product Features

    High resolution 15” color TFT display
    12 Channel waves.9 Parameters : ECG, Resp, NIBP, SpO2, TEMP,
    Light weight, compact and portable
    Built-in rechargeable lithium battery 
    Unique human voice alarm
    Rich screen layouts: Standard, Big font, Trend table, NIBP trend, ECG full lead, OxyCRG, IoC Layout
    Powerful storage: Maximum 720 Hour graphic and tabular trends of all
    Parameters, up to 10000 NIBP review ,200alarm review ,2Hours of all waveforms
    System software can be upgraded through network
    Data and waveforms color be adjustable
    Arrhythmia analysis and ST segment analysis
    Suitable for adult, paediatric and neonatal patient

    Options 
    Invasive Blood Pressure (IBP)
    Carbon dioxide ( CO2) Module
    Thermal Printer
    Touch Screen
    Three display output: TTL / CMOS LCD, LVDS LCD and standard VGA output

    Built-in Wifi (MD9015)

Product Specifications

    ECG 
    Lead Mode 5 Leads ( R, L, F, N, C or RA, LA, LL, RL, V)
    Lead selection I, II, III, avR, avL, avF, V, 
    Waveform 2 ch 
    Lead mode 3 Leads ( R, L, F or RA, LA, LL)
    Lead selection I, II, III, 
    Waveform 1 ch 
    Gain x2.5mm/mV, x5.0mm/mV, x10mm/mV, x20mm/mV, auto 
    HR and Alarm  
          Range 
          Adult 15 ~ 300 bpm
          Neo/Ped 15 ~ 350 bpm
          Accuracy ± 1% or ± 1bpm, which great
          Resolution 1 bpm
         Sensitivity > 200 (uV ) 
    Differential Input Impedance > 5 MΩ
    CMRR
          Monitor > 105 dB 
         Operation > 105 dB
         Diagnosis > 85 dB
    Electrode offset potential ±300mV
    Leakage Current < 10 uA
    Baseline Recovery < 3 S After Defi.
    ECG Signal Range ±8 m V ( Vp-p ) 
    Bandwidth    
         Surgery 1 ~ 15 Hz
         Monitor 0.5 ~ 35 Hz
         Diagnostic 0.05 ~ 100 Hz
    Calibration Signal 1 (mV ), Accuracy : ±5%  
    ST Segment Monitoring Range   
    Measure and Alarm -2.0 ~ +2.0 mV
    ARR Detecting  
    Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC 
        Alarm Available
        Review Available 

    RESPARATION 
    Method Impedance between RF (RA-LL)
    Differential Input Impedance >2.5 MΩ
    Measuring Impedance Range: 0.3~5.0Ω
    Base line Impedance Range: 0 – 2.5 KΩ
    Bandwidth 0.3 ~ 2.5 Hz
    Resp. Rate 
    Measuring and Alarm Range 
         Adult 0 ~ 120 rpm
         Neo/Ped 0 ~ 150 rpm 
    Resolution 1 rpm
    Accuracy ±2 rpm 
    Apean Alarm 10 ~ 40 S 

    NIBP 
    Method Oscillometric
    Mode Manual, Auto, STAT
    Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min)
    Measuring Period in STAT Mode 5 Min
    Pulse Rate Range 40 ~ 240 bpm
    Alarm Type SYS, DIA, MEAN
    Measuring and alarm range    
    Adult Mode
                 SYS 40 ~ 270 mmHg
                 DIA 10 ~ 215 mmHg
                 MEAN 20 ~ 235 mmHg 
    Pediatric Mode
                 SYS 40 ~ 200 mmHg
                 DIA 10 ~ 150 mmHg 
                 MEAN 20 ~ 165 mmHg
    Neonatal Mode 
                 SYS 40 ~ 135 mmHg
                 DIA 10 ~ 100 mmHg
                 MEAN 20 ~ 110 mmHg 
    Resolution 
                  Pressure 1mmHg
                  Accuracy
                  Pressure  
    Maximum Mean error ±5mmHg
    Maximum Standard deviation ±8mmHg 
    Overpressure Protection
                 Adult Mode 297±3 mmHg 
                 Pediatric Mode 240±3 mmHg
                 Neonatal Mode 147±3 mmHg 

    SpO2 
    Measuring Range 0 ~ 100 %
    Alarm Range 0 ~ 100 %
    Resolution 1 %
    Accuracy 70% ~ 100% ±2 %
                                         0% ~ 69% unspecified
    Actualization interval about 1 Sec.
    Alarm Delay 10 Sec. 
    Pulse Rate 
    Measuring and Alarm Range 0~254bpm
    Resolution 1bpm 
     Accuracy ±2bpm 

    TEMPERATURE 
    Channel 1
    Measuring and Alarm Range 0 ~ 50 C
    Resolution 0.1~C
    Accuracy ±0.1~C 
    Actualization interval about 1 Sec.
    Average Time Constant < 10 Sec. 

    IBP 
    Label ART, PA, CVP, RAP, LAP, ICP, P1, P2
    Measuring and alarm range 
        ART 0 ~ 300 mmHg
        PA -6 ~ 120 mmHg
        CVP/RAP/LAP/ICP -10 ~ 40 mmHg
        P1/P2 -10 ~ 300 mmHg     
    Press Sensor 
    Sensitivity 5 uV/V/mmHg 
       Impedance 300-3000Ω
       Resolution 1 mmHg 
       Accuracy ±2% or ±1mmHg, which great about 1 Sec. 
     

Related Products

Promoted Products

Meditech Equipment Co.,Ltd is part of Meditech Group. Product(s) described may not be licensed or available for sale in all countries. Sonotech, Sonovet, iSonic, FOs2pro, Dolphi, Defi, HeartRec,miniScan,Cardios,SpirOx,iBreath, Meditech and all corresponding design marks are trademarks of Meditech. The symbol indicates the trademark is registered. Patent and Trademark Office and certain other countries. All other names and marks mentioned are the trade names, trademarks or service marks of their respective owners. Please see the Instructions for Use for a complete listing of the indications, contraindications, warnings and precautions.