• Etco2 anesthesia agent analyzer used to measure EtCO2, FiCO2, RR, EtN2O, FiN2O, EtAA, FiAA.
  • Etco2 anesthesia agent analyzer used to measure EtCO2, FiCO2, RR, EtN2O, FiN2O, EtAA, FiAA.
  • Etco2 anesthesia agent analyzer used to measure EtCO2, FiCO2, RR, EtN2O, FiN2O, EtAA, FiAA.
Capnoxi® plus AG 
Handheld EtCO2 monitor Detecting gas: EtCO2, FiCO2, RR,EtN2O, FiN2O, EtAA, FiAA. High Precision.
anesthesia agent analyzer used to measure EtCO2, FiCO2, RR, EtN2O, FiN2O, EtAA, FiAA.

Product Features

    Monitoring SpO2, pulse rate, ETCO2 and respiratory rate;
    Update 255-hour historical data replay on PC through USB
    Equip Spo2 and Respiratory Status Analysis Software, comprehensive data analysis
    Suitable for clinical and emergency circumstance
    SpO2, SpO2+CO2 selectable
     
    Software Function
    Professional Data Analysis Software for CO2 Monitoring
    Connecting the palm monitor to PC through USB
    Storing Unlimited patient’s data
    Displaying ETCO2,SPO2 Trends Table
    Controlling PC printer, outputting trend graph, trend table report
    Recording and storage object’s respiration details in clinical
     

    Inside Pump Made by

Product Specifications

    Main Unit’s Environment Requirement
    Working:          Temperature: 5℃~50℃; 
                            Relative humidity: 0~95%;          
                            Atmosphere Pressure: 70.0KPa~106.0KPa
    Storage:          Temperature: 0℃~70℃; 
                           Relative humidity: 0~95%;
                            Atmosphere Pressure: 22.0KPa~120.0KPa 
     
     
    LED                             
    Patient Alarm Indicator: Two colors: Yellow and Red
    Sound Indicator           
    Loudspeaker:         Play alarm voices
    Interfaces                     
     
    Power:                    12VDC power socket x 1
     
    USB:                       MINI USB socket x 1
     
    Measurement Specification
    Principle:                 NDIR single beam optics
    Sampling Rate:       90mL/min, ±10mL/min
    Initialization Time:   Waveform displaying in 20 seconds
     
    Range                           
    CO2:                0~99 mmHg, 0~13 %
    N2O:                0~100 VOL%
    ISO:                 0~6VOL%
    ENF:                0~6VOL%
    SEV:                0~8VOL%
    RR:                 2~150 bpm
     
    Resolution                   
    CO2:               0~40 mmHg    ±2 mmHg
                           40 ~99 mmHg  ±5% of reading
     
    N2O:               0~100VOL% 
                           ±(2.0 vol% +5% of reading)
     
    ISO:               0~6VOL%   
                          ±(0.3 vol% +2% of reading)
     
    ENF:               0~6VOL%   
                           ±(0.3 vol% +2% of reading)
     
    SEV:               0~8VOL%   
                           ±(0.3 vol% +2% of reading)
     
    RR:                ±1 bpm
    Apnea Alarm Time:    20~60s
     
     MAC value defines 
     
    • 1.0MAC: under an atmospheric pressure condition, give a person or animal skin stimulation, 50% of people or animals not body dynamic response or escape reflection occurs, the inhalation anesthetics in alveolar concentration. 
    • To achieve 95% of people don’t respond to stimuli, the MAC value should reach to 1.3. 
    • When MAC value is 0.4,most patients will wake up.
     
    Anesthetic agents
    Enflurane:         1.68
    Isoflurane:         1.16
    Sevoflurane:     1.71
    Halothane:        0.75
    N2O:                 100%
    Notice                Desflurane’s MAC1.0 values differ with age
    Age:                  18-30   MAC 1.0  7.25%
    Age:                   31-65   MAC 1.0  6.0%
     
     
     
    Power Specification
    Input Voltage:       12V DC
    Input Current:       2.0 A    
     
    Physical Specification
    Main Unit
    Weight:               0.65Kg
    Dimension:         192mm x 106mm x 44mm
     
    Hardware Specification
    TFT Screen
    Type:                   Colorful TFT LCD
    Dimension:          5.0 inch
     
    Battery                          
    Quantity:             4
    Model:                Rechargeable lithium battery
    Voltage:              3.7 V
    Capacity:            2200mAh
    Working time:     10 hours
    Recharging time: 4 Hours
     

     

Related Accessories

Related Products

Promoted Products

Meditech Equipment Co.,Ltd is part of Meditech Group. Product(s) described may not be licensed or available for sale in all countries. Sonotech, Sonovet, iSonic, FOs2pro, Dolphi, Defi, HeartRec,miniScan,Cardios,SpirOx,iBreath, Meditech and all corresponding design marks are trademarks of Meditech. The symbol indicates the trademark is registered. Patent and Trademark Office and certain other countries. All other names and marks mentioned are the trade names, trademarks or service marks of their respective owners. Please see the Instructions for Use for a complete listing of the indications, contraindications, warnings and precautions.

Legal notice Terms and conditions Cookie policy Privacy Policy Professional organisations Careers

2024© Copyright احدى فروع شركة ميديتك جروب و المركز الرئيسي الإقليمي لمنطقتي الشرق الأوسط و شمال إفريقيا | Meditech Group (Middle East branch).

Meditech Group

To Top